Keeler KAT

Keeler

KAT

Mechanical Goldmann applanation tonometer

Fully mechanical Goldmann applanation tonometer read from a calibrated drum, so no battery is needed. Offered as R-type and T-type for H-style slit lamps and as the newer Z-type for Z-style (lower-illumination) slit lamps.

Serviced by manufacturer-trained engineers. New devices include two years of free service – only the engineer's travel is charged.

Highlights

  • Smooth drum control with readings from 0 to 80 mmHg
  • Calibration arm supplied for routine accuracy checks
  • Accepts re-usable Goldmann prisms and Tonoclear single-use prisms
  • Z-type available for Keeler slit lamps and for other manufacturers' Z-style slit lamps
  • British-made, hand-assembled in Windsor

Technical specifications

Measurement range
0–80 mmHg
Scale division
1.96 mN (2 mmHg)
Measurement uncertainty
≤ 0.49 mN up to 32.67 mN; ≤ 1.5 % above
Hysteresis
≤ 0.49 mN
Measuring force
lever weight, no battery
Versions
R-type (fixed) / T-type (removable) / Z-type (Z-style slit lamps)
Weight
0.82 kg (R) / 0.48 kg (T) / 0.76 kg (Z)
Standard
ISO 8612:2009
Medical device risk class
Class I

More in this category